Partenaires

CNRS Universite Grenoble Alpes UPS LaNEF INSA TOULOUSE EMFL NEXT

Home


Home page > Experiments and techniques > Set-up for large facilities (Synchrotron/Neutron) > Pulsed fields for synchrotron applications

Français   

Pulsed fields for synchrotron applications

Key parameters for magnet systems developed for synchrotron applications

X-ray absorption spectroscopy Powder diffraction Single crystal diffraction
Geometry Faraday Faraday Voigt
Coil solenoid solenoid conical bore ±31° split-pair
Maximum field 30 T 30 T 31 T
Pulse duration (total) 23 ms 23 ms 57 ms
Pulse duration (≥ 90% Bmax) 5.8 ms 5.8 ms 10.5 ms
Repetition rate at Bmax 10/hour 6/hour 2/hour
Sample diameter 9 mm 4 mm 3 mm
Sample temperature 1.5 -250 K 9 -250 K 1.5 -250 K


Members of the lab implied in this activity

J. Béard, J. Billette, F. Duc, P. Frings, W. Knafo, M. Nardone, J.P. Nicolin, A. Poux, G.L.J.A. Rikken, A. Zitouni


Former members of the lab implied in this activity

X. Fabrèges, K. Chesnel

X-ray diffraction in pulsed fields